NULL
K/J Type thermocouple input
Multi-channels scanner (T1-->T2-->T1 - T2)
Measurement Range
Type J : (-200 ºC to 760 ºC ) , (-328 ºF to 1400 ºF ) Type K: (-200 ºC to 1370 ºC ) , (-328 ºF to 2498 ºF )
Basic Accuracy : 23 +/- 5 ºC Calibration
Unit
Resd.
Range
Accuracy
Type K
Type J
ºC
0.1 ºC
-200 ºC ~0 ºC
+/-(0.2%rdg + 1 ºC )
0 ºC ~200 ºC
0 ºC ~150 ºC
+/-(0.1%rdg + 0.8 ºC )
1 ºC
200 ºC ~1370 ºC
150 ºC ~760 ºC
+/-(0.2%rdg + 2 ºC )
ºF
0.2 ºF
-328 ºF ~32 ºF
+/-(0.2%rdg + 2 ºF )
32 ºF ~392 ºF
32 ºF ~302 ºF
+/-(0.1%rdg + 1.6 ºF )
2 ºF
392 ºF ~2498 ºF
302 ºF ~1400 ºF
+/-(0.2%rdg + 3 ºF )
Power Source
One 9V battery. 006p or NEDA1604 or IEC6F22
Battery Life
Approx. 100hr
Operating / Storage Condition
0 ºC to 40 ºC (32 ºF to 104 ºF )below 80%RH. -10 ºC to 60 ºC (14 ºF to 140 ºF ) below 70%RH.
Dimensions
135(L) x 72(W) x 31(H)
Weight
Approx. 235g with battery
Accessories
Instruction Manual Holster ( option )
OPTIONAL TEMPERATURE PROBES K(CA) TYPE THERMOCOUPLE
Model
Testing Range
Description
TP-K01 Bead
-50 ºC ~200 ºC /-58 ºF ~392 ºF
for general condition, especially for complex and place hard to reach.
TP-K02 Immersion
-50 ºC ~1000 ºC C/-58 ºF ~1832 ºF
for temperature measurement of liquid or gels.
TP-K03 Surface
-50 ºC ~750 ºC /-58 ºF ~1382 ºF
for surface measurement.
No reviews yet. Be the first to review this product!